VNA Calibration
Calibration removes systematic errors from VNA measurements. Without calibration, your readings will include errors from cables, adapters, and the VNA itself.
When to Calibrate
- First use on a new frequency range
- After changing cables or adapters
- After significant temperature changes
- If measurements look wrong or inconsistent
What You Need
Calibration Standards
| Standard | Purpose |
|---|---|
| 50Ω Load | Reference impedance |
| Open | Open circuit (can use bare SMA) |
| Short | Short circuit |
| Through | Connects Port 1 to Port 2 |
Calibration Procedure
-
Set the sweep range
Ask Claude: “Set sweep from 144 to 148 MHz with 101 points”
Calibration is frequency-specific. Always calibrate at the frequencies you’ll measure.
-
Connect the LOAD
Connect the 50Ω load termination to Port 1.
Ask Claude: “Run cal load”
-
Connect the OPEN
Remove the load, leave Port 1 open (or use an open standard).
Ask Claude: “Run cal open”
-
Connect the SHORT
Connect the short standard to Port 1.
Ask Claude: “Run cal short”
-
Connect the THROUGH
Connect Port 1 to Port 2 with a short cable or barrel adapter.
Ask Claude: “Run cal thru”
-
Optional: Isolation
Connect 50Ω loads to both ports.
Ask Claude: “Run cal isoln”
This corrects for crosstalk between ports.
-
Apply calibration
Ask Claude: “Run cal done”
This computes and applies the error correction coefficients.
-
Save calibration
Ask Claude: “Save calibration to slot 0”
Saves to NanoVNA flash for later recall.
Verification
After calibration, verify it worked:
- Load test: Connect 50Ω load. S11 should be < -40 dB.
- Open test: Leave port open. S11 should be ~0 dB.
- Through test: Connect through. S21 should be ~0 dB, S11 < -40 dB.
Calibration Slots
The NanoVNA stores calibrations in flash memory:
| Model | Slots |
|---|---|
| NanoVNA-H | 0-4 |
| NanoVNA-H4 | 0-6 |
Each slot holds one calibration. Common approach:
- Slot 0: Full range (50 kHz - 900 MHz)
- Slot 1: HF (1.8 - 30 MHz)
- Slot 2: VHF (50 - 300 MHz)
- Slot 3: 2m band (144 - 148 MHz)
- Slot 4: 70cm band (420 - 450 MHz)
Recalling Calibration
Ask Claude: “Recall calibration from slot 0”
The NanoVNA will use the stored calibration. Note that calibration must match the current sweep settings.
Troubleshooting
High return loss with load
- Clean connectors with isopropyl alcohol
- Check for damaged center pin
- Try a different load
S21 shows high loss through
- Cable may be damaged
- Adapters may be loose
- Recalibrate with the exact cables you’ll use
Calibration doesn’t apply
- Frequency range must match
- Try recalibrating
- Check firmware version